Standalone monopolar high frequency electrosurgery unit for soft tissue surgery with:
-
Maximum output power – cutting mode: 50 W at non-inductive load impedance of 500 ohms
-
Maximum output power – coagulation mode: 30 W at non-inductive load impedance of 500 ohms
-
Automatic power-tuning system based on the concept of impedance matching
-
Work frequency: 27.126 MHz yielding, no neutral electrode necessary
Operation:
-
On/off control in handpiece
-
Standby and choice between cut and coagulation via touch buttons on the control unit
-
Indication of standby, cut or coagulation by LEDs on the control box
Asepsis:
Including:
-
Handpiece cable (black)
-
Handpiece holder
-
Standard set of electrodes with both straight and loop-shaped electrodes for facial and intraoral surgery, and ball electrodes for coagulation in casing: 220501, 220504, 220510, 220511, 220513, 220514 and 220515
Dimensions of the control box (W x D x H): 17,5 x 24,8 x 5 cm
Packed in box (W x D x H): 40,5 x 20,5 x 11,5 cm
Weight of box: 3,7 kg
Contraindications:
HF soft tissue handling is contraindicated in patients with cardiac pacemakers. Medical consultation should always be considered prior to using HF soft tissue handling in those patients who may be systemically medically compromised, especially those with abnormal healing processes such as diabetes, blood dyscrasias, collagen disturbances and patients undergoing irradiation treatment.
XO ODONTOSURGE, 230V (OSU430)